Error
Device behaviour Cause
13
Error screen
Switch to normal
mode after 10
seconds
22, 23
Error screen
Switch to normal
mode after 10
seconds
-
The device does
not turn ON,
indicator LED is
RED
-
The device does
not turn ON,
indicator LED is
blinking RED
-
LED Wireless
charger blinking
blue.
-
Device not
functioning
Table 1 error number matrix

Other indications/faults

Output indicator
When the output indicator lights up yellow, the
related patient-output is active. The intensity-level
is set higher than 0.0 mA, so the output delivers an
electrical signal.
i-Pulz 2
Memory error
Device was
not turned off
properly last time.
When noticed
multiple times
contact the
distributor
The battery is too
low
Internal error with
battery
The charger has
become too
hot. Or ambient
tempera-ture
is below 0
degrees Celsius
(32 degrees
Fahrenheit).
What to do
Press button, device ready
to use. Log-fi le is cleared.
Press the button to use the
device.
The internal time of the
device is reset; this can
give a misrepresentation
in de compliance-menu.
Set the time in the settings
menu.
Put the device on the
charger
Remove the i-Pulz 2 to cool
down the charger.
Remove the i-Pulz 2 to cool
down the charger.
Reset the device by
holding the Power button
for 5 seconds.