Operation - Thermo Scientific ADVANCED Series Quick Start Manuals

Hide thumbs Also See for ADVANCED Series:
Table of Contents

Advertisement

Operation

Section 4
ADVANCED
The Thermo Scientific ADVANCED Series of Heated Immersion Circula-
tors have a digital display and easy-to-use touch pad, five programmable
Heated
setpoint temperatures, acoustic and optical alarms. All circulators offer adjust-
Immersion
able high temperature protection.
Circulator
145
110
180
75
215
40
°C
This label indicates read the instruction manual before starting the circulator.
Use this button to place the circulator in and out of standby, see page 4-3 for
more details. The blue LED illuminates when standby is enabled.
Use these navigation arrows to move through the circulator displays and to
adjust values.
Pressing this button once to make changes on the immersion circulator's dis-
play screen. In most cases, pressing it again is required to save the change.
Use this button to cancel any changes and to return the immersion circula-
tor to its previous display. Canceling a change can only be made before the
change is saved. In some cases, it is also used to save changes.
Note Holding this button for five seconds resets the display contrast to the
default level and also brings up the language menu to change, if needed, the
Settings-Display Options
displayed language. See
in this Section.
Used for adjusting and resetting the High Temperature Cutout. Details are
explained in this Section.
°C
4-1
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents