Section 4 Operation
Settings - Application Settings
Real Temperature Adjustments (
speed, configure the interface (optional), set the clock, turn the timer on or off, and turn auto restart and
energy savings on or off.
Application Settings
1. With
press
to view:
Setpoints
Alarms
Fluid Type
Pump Adjustment
Menu
Setpoints
3. With
highlighted, press
to display the list. Use the up/down arrows to
SP
highlight the desired
SP5
to display
.)
SP1
SP2
SP3
SP4
Menu
If this temperature on the Start/Status Displays does not accurately reflect the actual temperature in the
reservoir, an RTA can be applied. The RTA can be set ±10°C (±18°F).
As an example, if the circulator's temperature is stabilized and displaying 20°C but a calibrated reference
thermometer reads 20.5°C, set the RTA to -0.5°C. After you enter a RTA value allow the temperature to
stabilize before verifying the temperature in the bath. Note If display accuracy is required, we recom-
mend repeating this procedure at various setpoint temperatures and on a regular basis.
Note You cannot adjust the setpoint closer than 0.1°C to either of the fluid's system limits, see Fluids
Type in this Section.
5. With the desired line highlighted press
The right-most digit will have a cursor beneath it.
Use the left and right arrows to move the cursor
to the desired digit and then use the up and down
arrows to change the value. Once all the desired
changes are made, press
or
to cancel it.
4-8
is used to view/adjust the circulator's five Setpoints (
RTA
), enable/disable the alarms, change the fluid type, set the pump
highlighted
. (Use the down arrow
.
to save the change
2. Scroll down for additional options.
Set Clock
On/Off Timer
Auto Restart
Energy Savings
Menu
4. Press
.
The Setpoint and RTA are changed using the
same procedure. With the desired setpoint high-
lighted press
to display the submenu.
SP1
RTA int
RTA ext
Menu
35.0°C
SP
) and
xx.x
xx.x
xx.x
Thermo Scientific