Section 4 Operation
Setup
CAUTION
CAUTION
Start Up
4-2
Leave refrigerated baths in an upright position at room temper-
ature (~25°C) for 24 hours before starting. This will ensure the
lubrication oil has drained back into the compressor.
Before starting, double check all USB (optional), electrical and
plumbing connections.
D o not run the circulator until fluid is added to the reservoir. Have
extra fluid on hand. If the circulator does not start refer to Section 5
Troubleshooting.
• On refrigerated baths, place the circuit protector
located on the rear to the I position.
• Place the circuit protector located on the back of the
immersion circulator to the I position.
After a slight delay the blue LED on the front panel illuminates.
• Press
goes out.
• Ensure the start symbol has a highlight box around it, if not
use the arrow keys to navigate to the symbol.
20.0°C
Highlighted
Start Symbol
• Press
into a stop symbol (
20.0°C
Stop Symbol
Note The pump starts immediately but refrigerated baths take up to
30 seconds before the compressor starts.
Note After start up, check all the plumbing connections for leaks.
, the Start Display appears and the blue LED
SP1
Menu
24.2°C
Start Display
. The circulator starts and the start symbol turns
).
Reservoir
In
Fluid
Temperature
Thermo Scientific
I
O