5. Control
The electrochemical sensors for gas analysis are exposed to a process of
wear and tear and they do age. In the course of their use, they change their
initial values depending on the gas concentration, the duration of exposure
to gas and the degree of contamination of the measurement gas.
The device monitors the sensors and corrects any deviation. However, if
these deviations and thus the measurement error to be expected become
too large, a message is displayed. In this case the measuring module has to
be exchanged.
The current status values for the sensors are displayed in the "Control"
menu. The cursor keys can be used to scroll through the 4 control screens.
ecom GmbH
O2 19744 mV
Am Großen Teich 2
CO
58540 Iserlohn
Batt 8.09 V
Operation hours:
8.45 hrs
Total
: 18.75 hrs
Next unit check: 01.07.22
Service tel.
:0800-945-5000
Further Pages:
Programmversion :V1.14 08.06.21
Serial no.
:ST-0001
CO purges
: 15
Error counter
: 21
Operation hours:
8.45 hrs
Total
: 18.75 hrs
Next unit check: 01.07.22
Service tel.
:0800-945-5000
Further Pages:
ecom-ST 1.0
Indication
Sensor #### mV
7 mV
Battery #,## V
Operation hours
Total operation hrs
Next maintenance
Service tel.
Program version
Serial number
CO purges
Error counter
Meaning
Output voltage of
sensor in mV
Output voltage of
battery
Operation hours since
last service
Total operation hours
Date of next
maintenance
Phone number of next
service centre
Version of instrument
firmware
Serial number of the
instrument
Number of active CO
switch-offs until now
Number of error
messages displayed
since then
27