OPERATING INSTRUCTIONS
Version
ecom-ST 1.0
State July 2021

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ST and is the answer not in the manual?

Questions and answers

Summary of Contents for Ecom ST

  • Page 1 English OPERATING INSTRUCTIONS Version ecom-ST 1.0 State July 2021...
  • Page 2: Table Of Contents

    2.1 Wall-mounting protective housing (option) 2.2 Rack installation 3. Starting Up 4. Measurement Procedure 4.1 Preparation 4.2 Flue gas analysis 4.3 Measurement screens 4.4 Draught measurement 4.5 CO overload protection 5. Control 6. Adjustments 6.1 Remote Setup 6.1.1 Modbus ecom-ST 1.0...
  • Page 3 6.1.2 Time control 6.1.3 Configuring the Analog outputs (option) 7. Data Processing 7.1 Data logger 7.2 Data connection to the PC 7.2.1 ecom-DAS NT2 7.2.2 ecom-DAS 5 7.3 ecomModbus 8. Maintenance 8.1 Fine dust filter 8.2 Gas sensors 8.3 Calibration 8.4 Removal / re-installation of modules...
  • Page 4: Intended Use

    Intended Use The ecom-ST is a stationary measuring instrument made by ecom GmbH, which is intended for the semi-continuous exhaust gas measurement at combustion systems and which is permanently installed on site. The device is operated either by the internal time control or by time control via Modbus.
  • Page 5: General Safety Information

    Damage to the sensors Substances containing solvents damage the device's sensors. Avoid contact with substances containing solvents! CAUTION Damage to electronic components Electrostatic discharge can damage electronic components. Before touching the inside of the housing, touch a grounded component! ecom-ST 1.0...
  • Page 6: Device Structure

    NOTE Check fine dust filter and PTFE filter regularly! Change filter when the grey staining equals <3 on the soot dot comparison scale! 1.Device Structure 1.1 Components Device with protective housing (option) ecom-ST 1.0...
  • Page 7 Graphic display Condensation pump for CO sensor Gas cooler with Fresh air inlet PTFE filter fine dust filter ecom-ST 1.0...
  • Page 8: Connections

    1.2 Connections Measuring gas USB connection Analog outputs (option) outlet Mains power Slot for multi Condensation connection - max. media card (SD) outlet cable length 3 m Network Mains supply connection ventilation fan ecom-ST 1.0...
  • Page 9: Keypad

    Battery LED (for Enter key (confirm indication of ON/OFF key selection) charging status) Function keys Cursor control (assignment shown (up/down/right/left) on display) NOTE In input mode, the numerical function of the keys is used to enter figures. ecom-ST 1.0...
  • Page 10: Interfaces

    PC or laptop. Data can be recorded using the ecom PC software DAS NT 2 or DAS5. A driver is required for communication between the device and the ecom PC software DAS NT 2 or DAS5. PC software and USB drivers are...
  • Page 11 The Modbus slave address is set to 0x2 in the delivery state. The maximum cable length is 50 m. If the ecom-ST is used as a terminal device, a termination of 120  is provided. The termination is active on delivery.
  • Page 12 The pin assignment is listed below. Pin assignment 8-pin socket 8-pin plug Position at device The following transmission parameters must be set at the ecom-ST ("Adjustments" / "Internal" / "COM module"): • 9600 Baud (recommended adjustment) • Protocol Modbus RTU ecom-ST 1.0...
  • Page 13: Ethernet

    1.4.3 Ethernet An Ethernet interface is available for connecting the device via Modbus TCP. The ecom-ST can be integrated into a network via the LAN connection on the front. The following transmission parameters must be set at the ecom-ST ("Adjustments" / "Internal" / "COM-Modul"): •...
  • Page 14 You can set the password for the administrator and for the creation of additional users under the menu item "User". ecom-ST 1.0...
  • Page 15 To adapt the network configuration, you can change the network settings under "Network" by selecting "eth0" → "Configuration". ecom-ST 1.0...
  • Page 16: Analog Outputs (Option)

    Signal channel 7 GND channel 5 Signal channel 8 GND channel 6 Signal channel 5 GND channel 7 Signal channel 6 Each channel can be assigned a measurement or calculation variable with free scaling (see chapter “Configuring the Analog outputs”). ecom-ST 1.0...
  • Page 17: Unheated Gas Sampling System

    1.6 Unheated gas sampling system The ecom-ST can be equipped with a so-called NO tubing. The NO tubing effectively minimizes the washout losses of water-soluble gas components like NO and SO . Thanks to its smooth surface and the increased flow rate, the condensation formation and the correlated washout impact are minimized.
  • Page 18: Data Storage

    Loss of data and damage to the data carrier are possible. Do not pull out cards while data is being recorded! All values of a recording are saved in a csv file with the name " ST_DL- xx.csv ". Successive files are automatically numbered from 00. ecom-ST 1.0...
  • Page 19: Installation

    2. Installation NOTE Fastening material is not included and must be selected on site according to the individual circumstances. 2.1 Wall-mounting protective housing (option) Dimensional drawing ecom-ST 1.0...
  • Page 20: Rack Installation

    2. Depending on the wall and the individual conditions on site, screw protective housing to the wall. NOTE Do not use the housing as a storage space! 2.2 Rack installation NOTE 6 HE must be complied with! ecom-ST 1.0...
  • Page 21: Starting Up

    In the main menu 4 submenus are Control displayed. Remote setup Menu point Meaning Gas analysis Perform flue gas analysis Settings Modify instrument settings Control Check operation state of instrument Remote Setup Adjust data transfer ecom-ST 1.0...
  • Page 22: Flue Gas Analysis

    If a memory card is inserted, you are asked for internal data processing, which must be answered with <F4> = no (internal data processing with the ecom-ST is irrelevant). Then the suitable fuel must be selected from a table.
  • Page 23: Measurement Screens

    With <F1> you can switch from the measured value display to a menu item or a function selected previously (see chapter “Adjustments”). Possible settings are: Fuel type, Eff. (C), Internal, Adjustments, Standby, Purge. You can also switch to the measured value display from any menu item while pressing <F1>. ecom-ST 1.0...
  • Page 24 3. Select the desired parameter with the cursor keys (right / left). 4. Repeat the procedure until the change is completed. 5. Deactivate the function with <Enter>. The measured and calculated values in measurement screen 4 can be set with the cursor keys (right / left). ecom-ST 1.0...
  • Page 25: Draught Measurement

    Disconnect the draught hose on the Recorded value: device and press <F4>. The sensor is -0.12 hPa recalibrated. Reattach the draught hose. The display shows the exact measured value, that can be frozen with the <Store> key. ecom-ST 1.0...
  • Page 26: Co Overload Protection

    The CO display in the status bar of the measurement screen no longer flashes and the status LED <Info> changes from "red" to "green". ecom-ST 1.0...
  • Page 27: Control

    : 18.75 hrs firmware Next unit check: 01.07.22 Serial number Serial number of the Service tel. :0800-945-5000 Further Pages: instrument CO purges Number of active CO switch-offs until now Error counter Number of error messages displayed since then ecom-ST 1.0...
  • Page 28 205 hrs (history) of the device are listed: 23.08.2020 322 hrs date of maintenance, --.--.-- --.--.-- operating hours at the Further Pages: maintenance time. Sensor information Information as to age and calibration data of the sensors. Further Pages: ecom-ST 1.0...
  • Page 29: Adjustments

    Change of the set fuel type Country-specific Clock set Adjustment of internal clock Sub-menus for further Internal adjustments -- see following table -- NOTE The units mg/kWh and mg/MJ are always calculated and displayed with a reference oxygen content = 0 %. ecom-ST 1.0...
  • Page 30 All settings in the table Settings for the USB interface on the following page All settings in the table COM module Settings for the COM interface on the following page ecom-ST 1.0...
  • Page 31 Transfer speed Enhanced = DAS 5 DAS = DAS NT 2 Protocol Modbus RTU Modbus / TCP ---- = none HINWEIS Only one Modbus application can be active, either Modbus RTU via RS485 or Modbus / TCP via Ethernet. ecom-ST 1.0...
  • Page 32: Remote Setup

    Indication Meaning Example TxD buffer Received data See ecomModbus RxD buffer Sent data See ecomModbus Device address ecom-ST device address Standard setting 2 Setup Parameter Meaning Setting change Change of the ecom-ST device Entry after activation Device address address for Modbus RTU...
  • Page 33 Nitrogen monoxide content (adjusted 2 unit) Nitrogen dioxide content (adjusted 2 unit) Sulphur dioxide content (adjusted 2 unit) Total nitrogen oxide content in ppm Total nitrogen oxide content (adjusted unit) Total nitrogen oxide content (adjusted 2 unit) ecom-ST 1.0...
  • Page 34: Time Control

    8. Input 20 mA value via numerical keys. 2 T.Gas 20.0 400.0 °C 9. Confirm with Enter. 12 CO2 0.00 20.00 % 4 O2 0.00 5.00 % 5 CO 500 ppm 6 NO 100.0 ppm 7 NO2 10.0 ppm 28 NOx 100.0 ppm ecom-ST 1.0...
  • Page 35 9 mA Fresh air purging active 10 mA Device in measuring mode, but the CO sensor is switched off 20 mA Device in measuring mode HINWEIS Valid measured values are only available if the status is 20 mA. ecom-ST 1.0...
  • Page 36: Data Processing

    SD card in which all subsequent measured values of the recording are saved. The files have a predefined name and are numbered consecutively. Example: • ST_DL-00.csv → oldest file • ST_DL-01.csv • ST_DL-02.csv • ST_DL-xx.csv → latest file ecom-ST 1.0...
  • Page 37: Data Connection To The Pc

    7.2.1 ecom-DAS NT2 In addition to the data logger recording, the data can also be transferred directly to the PC software "ecom-DAS NT 2" while the measuring operation is running. The USB interfaces can be used for communication with the software. The...
  • Page 38: Ecom-Das

    7.2.2 ecom-DAS 5 The software "ecom-DAS 5" can be used for the display and logging of measured values and measured value progressions. Measurement data can be exchanged between the software and the device via the SD card. A recording of point measurements and course measurements during the measuring operation can be implemented via USB.
  • Page 39: Maintenance

    Only use sensors or probes provided by ecom GmbH or from an ecom-authorized service center! NOTE If service work is not carried out by ecom GmbH or a service center authorized by ecom GmbH, the warranty for the device expires with immediate validity! 8.1 Fine dust filter...
  • Page 40: Gas Sensors

    NOTE The O sensor must display a value >7000 mV. If this is not the case, the sensor must be replaced. Contact the ecom service centre next to you! NOTE If an error message occurs during the calibration phase, a further calibration phase can help to calibrate the sensors sufficiently.
  • Page 41: Calibration

    -CO / NO / NO2 / SO2 calibration (Important: Span gas must get pressure-free to the instrument!) -Flow with span gas -Wait for stabilisation of value Span gas used by ecom (appoximate value): -By drift press <Enter> -Set correct value with number keys...
  • Page 42: Removal / Re-Installation Of Modules

    2. Loosen the 4 screws. 3. Pull the gas processing module out of the device. Reinstallation 1. Insert gas processing module into the device. 2. Tighten the 4 screws. Slip the hoses over the turned parts. Hoses Screw ecom-ST 1.0...
  • Page 43: Probe/Tubing

    The hot flue gas probe can damage the device or the transport system if it comes into contact with. Let the probe cool down before storing! 8.6 SBK system (option) The PTFE filter in the probe head should be changed regularly. ecom-ST 1.0...
  • Page 44: Technical Data

    300 hPa − 1100 hPa Air pressure ±2 hPa EC = Electrochemical sensor IR = Infrared sensor K = Catalytic Sensor T = Thermocouple NiCr-Ni PT = PT 2000 D = DMS bridge MV* = Measured value (higher value prevails) ecom-ST 1.0...
  • Page 45 Weight Approx. 8,6 kg Protective housing (option) Dim. (W x H x D) Approx. 600 mm x 350 mm x 265 mm Weight ecom-ST with protective housing approx. 25 kg NOTE Subject to technical changes! (stand July 2021) ecom-ST 1.0...
  • Page 46: Disposal

    • Operation hours etc. The instrument shows the error The sensor module must be message "O2 sensor mV too low" replaced. The instrument shows the error The sensor module must be message "xx sensor mV too high" replaced. ecom-ST 1.0...
  • Page 47 GmbH Am Großen Teich 2 D-58640 Iserlohn info@ecom.de www.ecom.de Telefon: +49 2371 945-5 Fax: +49 2371 40305 ecom-ST 1.0...