Kodak DirectView CR 825 Theory Manual page 70

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

THEORY GUIDE
10DEC05
PLATE POSITIONING AY
TG4825-1
Page
70 of 120
UPPER ARM
LOWER ARM
PLATE
POSITIONING
AY
EXTRACTION
BAR
LEAD SCREW
The LEAD SCREW is connected to the MOTOR SHAFT. When the LEAD SCREW rotates, it
moves the EXTRACTION BAR up and down.
The ARMS of the PLATE POSITIONING AY
keep the PLATE in the correct vertical
position for removal from the CASSETTE,
for scanning, and for inserting the PLATE
into the CASSETTE SHELL after scanning.
The PLATE POSITIONING AY has 2 ARMS
that each include:
LOWER ARM - when the EXTRACTION
BAR moves down to the position
immediately before scanning starts, the
LOWER ARM moves forward to touch
the back of the PLATE, which is partially
out of the CASSETTE SHELL. The
LOWER ARM keeps the PLATE
SCREEN from touching the WALLS of
the CASSETTE when it moves out of the
PLATE
CASSETTE.
UPPER ARM - after the LOWER ARM
moves forward to touch the back of the
PLATE, the UPPER ARM also moves
forward. It keeps the larger PLATES
H194_5013GCB
steady during scanning and when they
H194_5013GC
move out of and back into the
CASSETTE.