THEORY GUIDE
10DEC05
UNINTERRUPTIBLE POWER SUPPLY (UPS)
TG4825-1
Page
The UNINTERRUPTIBLE POWER SUPPLY UPS1 starts a system shutdown if the power is
111 of 120
not available for more than 15 seconds. If the power starts again before the 15 seconds, no
shutdown occurs and the system does not display any message to the user. If power is not
available for 15 seconds or more, the CR 825/850 SYSTEM is deactuated 105 seconds after
the power failure occurred. During the 105 seconds, the following actions occur.
The application software completes all actions, including shutdown of the operating system
of the INTERNAL PC.
Within 30 seconds of the first power failure, the system provides a window with a
message to the user about a shutdown that is about to occur.
The system completes transmission of any images sent at the time of the power failure.
If a CASSETTE was detected at the load position before starting the shutdown, the
system completes the scan, stores the raw image data on the hard disk, and erases the
SCREEN before the shutdown of the UPS.
The WEB BROWSER is the first software component to close down to prevent entering
any more PEC records or asking to view images.
Note
If aCASSETTE is placed in the load position after the shutdown process of the CR 825/850
SYSTEM has started, it is not scanned. No error message displays to the user because the
shutdown message has displayed on the TOUCH SCREEN MONITOR and it cannot receive
messages after that time.
When power is restored to the CR 825/850 SYSTEM, the operator can energize the
CR 825/850 SYSTEM using the normal procedure.