10DEC05
The CR 825/850 SYSTEM uses SCREENS with the following resolutions.
TG4825-1
Page
35 of 120
Resolution of the
SCREEN
General Purpose
(GP-25)
High Resolution
(HR)
Enhanced High
Resolution (EHR)
The light intensity of the exposed GP-25 SCREEN is less than the HR and EHR SCREENS.
The CR 825/850 SYSTEM adjusts for the difference in light intensity by adjusting the pixel
code values in the image processing.
thicker PHOSPHOR COATING than the HR SCREEN, making the
image less sharp
no special exposure procedures are necessary
thinner PHOSPHOR COATING than the GP-25 SCREEN, making
the image sharper
approximately 2 times the X-ray exposure is necessary
thinner PHOSPHOR COATING than the HR SCREEN, improving the
image quality
approximately 4 times the X-ray exposure
STORAGE PHOSPHOR CASSETTE
Description

Advertisement