THEORY GUIDE
10DEC05
The FRONT DOOR INTERLOCK SWITCH controls RELAYS on the MSC BOARD A1 that de-
TG4825-1
Page
energize the following systems.
110 of 120
Systems Protected by
INTERLOCK SWITCH
MSC BOARD A1
SLOW SCAN
CONTROLLER
BOARD A6
LASER DRIVER
PRE-REGULATOR
BOARD A18
LASER DIODE
DRIVER BOARD
A17
ERASE LAMPS
•
has an LED that indicates the status of the INTERLOCK SWITCH -
DS25
•
has LED INDICATORS for the 4 INTERLOCK RELAYS:
–
DS8
–
DS14
–
DS15
–
DS16
•
controls the INTERLOCK SWITCH for all DC MOTORS
•
+5 V DC is continually energized when the CR 825/850 SYSTEM is
energized
•
+24 V DC is de-energized when the INTERLOCK SWITCH is
opened
•
–24 V DC to the BOARD is de-energized when the INTERLOCK
SWITCH is opened
•
–24 V DC to this BOARD is routed through the GALVO BOARD, but
is not used by the GALVO BOARD
Note
The GALVO BOARD is continually energized when the CR 825/850
SYSTEM is energized.
6.5 V. DC from the LASER DRIVER PRE-REGULATOR BOARD A18 is
de-energized when the INTERLOCK SWITCH is opened.
120 V AC is de-energized when the INTERLOCK SWITCH is opened.
Description
Power Distribution