Kodak DirectView CR 825 Theory Manual page 107

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

THEORY GUIDE
10DEC05
Power Distribution
TG4825-1
Page
107 of 120
AC SOURCE
100-240V AC
50/60 Hz
H194_5023HC
INTERLOCK
SWITCH S8
TRANSFORMER T1 The TRANSFORMER T1 receives from 100 to 240 V AC, 50/60 Hz,
= INTERLOCK SWITCH
T1
TRANSFORMER
120V AC
50/60 Hz
Supply
The INTERLOCK SWITCH S8 controls RELAYS on the MSC BOARD,
which controls the 24 V DC power to the DC components, including
the MOTORS and SOLENOIDS. The INTERLOCK SWITCH
deactuates when the FRONT DOOR is opened.
and changes this power to 120 V AC, 50/60 Hz.
120V AC
UPS1
120V AC
120V AC
K1
RELAY
INTERNAL
PC
PS1
A1
MSC
POWER
BOARD
SUPPLY
A5
V1
24V DC
A6
GND
V2
12V DC
V3
-24 V DC
-24 V DC
ERASE
LAMPS
-6.5V DC
Description
Power Distribution
A2
A3
MCPU
DIGITIZER
BOARD
BOARD
5V DC
24V DC
PMT/DAS
BOARD
SLOW
SCAN
CONTROL
BOARD
DS1
TOUCHSCREEN
DISPLAY
A4
GALVO
BOARD
A18
LASER DRIVER
PRE-REGULATOR
BOARD
A17
LASER DIODE
DRIVER BOARD

Advertisement

Table of Contents
loading

This manual is also suitable for:

Directview cr 85056344825

Table of Contents