59 of 120
DIGITIZER BOARD
A3
GALVO
GALVO BOARD A4
Controls the operation of the GALVO BOARD, which moves the laser
beam.
Includes:
MOTOR - rotates the SHAFT
SHAFT - has a MIRROR at one end to send the laser beam
toward the F-THETA LENS
Includes a feedback circuit that controls the position of the MIRROR.
The position of the MIRROR determines the position of the laser
beam. The GALVO BOARD uses the following information to define
the desired position of the MIRROR:
clock signal from the DIGITIZER BOARD, which moves through
the desired position between the offset and amplitude.
size of the PLATE from the DIGITIZER BOARD, which receives
the information from the BAR CODE LABEL on the CASSETTE
values for the PLATE size, which were set up during calibration:
offset - starting point
amplitude - the distance to move across the SCREEN
Information from POSITION SENSORS determine the position of the
GALVO SHAFT in the scan - the actual position. The actual position
is compared to the desired position and corrections in position are
made to provide a smooth motion of the beam.
The GALVO BOARD also:
emits "Line Start" signals to the DIGITIZER BOARD to provide the
information that a line is complete and it can start another line
energizes the LASER DIODE DRIVER BOARD at times
determined by software