THEORY GUIDE
10DEC05
Overview of CR Technology
TG4825-1
Page
Operations
17 of 120
The following operations are necessary to capture the latent image in the STORAGE
PHOSPHOR SCREEN and change it to a digital image that can be viewed on a computer
screen and sent to a PRINTER.
•
Exposing the STORAGE PHOSPHOR SCREEN
•
Stimulating the PHOSPHOR
•
Changing Light Energy to an Analog Signal
•
Changing Analog Signals to Digital Signals
•
Processing the Digital Image
Exposing the STORAGE PHOSPHOR SCREEN
X-RAY
TUBE
H194_5033BC
latent
aerial
image
image
STORAGE PHOSPHOR SCREEN
Charged storage phosphors
proportional to X-ray energy
absorbed by screen.
Radiography Theory
Lighter values indicate that more
x-rays were absorbed by the
SCREEN - bone tissue
Mid-range values indicate that fewer
x-rays were absorbed by the
SCREEN - soft tissue
Darker values indicate that most
x-rays were absorbed by the
SCREEN - did not pass through the body