Calibration - E+E Elektronik EE29 Series Manual

Humidity/temperature transmitter
Table of Contents

Advertisement

i
1-point calibration Humidity:
i
Configuration software
5.4

Calibration

In addition to the manual calibration procedure on the circuit board (see Hardware,
chapter 7 "Humidity/Temperature Calibration"), new calibrations can be
performed using the EE29/31 software.
Note: A two-point calibration for temperature is only possible on the circuit board
and is not supported by the software.
Fast and easy calibration for accurate measurement results at a defined working
point (humidity point).
For calibration procedure see Hardware, chapter 7. "Humidity/Temperature Calibration"
1) Stabilise the probe of the desired humidity for min. 30 minutes.
2) Click on the Humidity "1-point calibration" button. The measured values will
now appear in both input fields.
3) Replace the value in the input field "Humidity Reading" with the reference
humidity (value of the saline solution or display of HUMOR 20).
4) By clicking on "Save", the humidity reading for the transmitter will be adjusted
to the reference humidity.
32

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ee31 seriesEe31-aEe31-fEe31-bEe31-dEe31-e

Table of Contents