RH / T adjustment and calibration

Humidity calibration and adjustment
Depending on the application and the requirements of certain industries, there might arise the need for
periodical humidity calibration (comparison with a reference) or adjustment (bringing the device in line
with a reference).
Calibration and adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For infor-
mation on the E+E capabilities in ISO or accredited calibration please see
www.eplusecal.com.
Calibration and adjustment by the user
Depending on the level of accuracy required, the humidity reference can be:
• Humidity Calibrator (e.g. Humor 20), please see www.epluse.com/humor20.
• Handheld device (e.g. Omniport30), please see www.epluse.com/omniport30.
• Humidity calibration Kit (e.g. E+E Humidity Standards), please see www.epluse.com/EE360.
Perform 1 or 2 point adjustment via EE-PCS Product Configuration Software or via display
(details see below).
Menu
Customer
adjustment
Rel. humidity
Temperature
Fig. 33
Adjustment menu
Fig. 34
Offset adjustment
Operating instructions for EE360 Moisture in Oil Transmitter
Offset
2 point adjustment
i
Calibration status
Reset to factory calibration
reference value
measurement value
Offset adjustment point
value
Reference value
Current
measured value
Reference
Reference value
value low
Reference value
Current measured
high
value
Measured value