Rh / T Adjustment And Calibration - E+E Elektronik EE310 Operating Manual

Humidity/temperature transmitter
Hide thumbs Also See for EE310:
Table of Contents

Advertisement

6.4

RH / T adjustment and calibration

Humidity calibration and adjustment
Depending on the application and the requirements of certain industries, there might arise the
need for periodical humidity calibration (comparison with a reference) or adjustment (bringing the
device in line with a reference).
Calibration and adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For
information on the E+E capabilities in ISO or accredited calibration please see
www.eplusecal.com.
Calibration and adjustment by the user
Depending on the level of accuracy required, the humidity reference can be:
• Humidity Calibrator (e.g. Humor 20), please see www.epluse.com/humor20.
• Handheld device (e.g. Omniport30), please see www.epluse.com/omniport30.
• Humidity calibration Kit (e.g. E+E Humidity Standards), please see www.epluse.com/EE310.
Perform 1 or 2 point adjustment via EE-PCS Product Configuration Software or via display
(details see below).
Menu
Customer
adjustment
Rel. humidity
Temperature
Fig. 27
Adjustment menu
Fig. 28
Offset adjustment
26
Offset
2 point adjustment
i
Calibration status
Reset to factory calibration
reference value
measurement value
Offset adjustment point
value
Reference value
Current
measured value
Reference
Reference value
value low
Reference value
Current measured
high
value
Measured value
Operating instructions for EE310 Humidity / Temperature Transmitter