9.2.3

Analog signal flow OUT2

Select [SEL2] and set FLOW
Select [ou2] and set analog signal:
I (4 to 20 mA)
Select [ASP2] and set the value at which 4 mA
is provided.
Select [AEP2] and set the value at which 20 mA
is provided.
9.3
Settings for consumption
quantity monitoring
9.3.1
Quantity monitoring by pulse signal OUT1
and OUT2
Select [SELx] and set FLOW
Select [oux] and adjust the pulse output: ImP
Select [ImPSx] and set the pulse value (= flow quantity
at which a pulse is provided):
1.
Press [▲] or [▼] to select the setting range.
2.
Press [●] briefly to confirm the setting range
3.
Press [▲] or [▼] to set the required numerical
value.
4.
Press [●] briefly to apply the value.
Select [ImPRx] and set YES.
Menu OUT2:
[SEL2]
[ou2]
[ASP2]
[AEP2]
Menu OUTx:
[SELx]
[oux]
[ImPSx]
[ImPRx]