Unlocking:
Make sure that the instrument is in normal operating mode.
▪
▪
Press [▲] and [▼] simultaneously for 10 seconds until [ Reset menu
lock] is displayed
9.1.4
Timeout
If no key is pressed for 30 seconds while setting a parameter, the
instrument returns to the operating mode with the value unchanged.
9.2
Settings for flow monitoring
9.2.1
Limit value monitoring OUT1 or OUT2 /
hysteresis function
▪
Select [SELx] and set FLOW
Select [oux] and adjust the switching signal:
▪
-
Hno = hysteresis function / normally open
-
Hnc = hysteresis function / normally closed
▪
Select [SPx] and set the value at which the output
switches.
▪
Select [rPx] and set the value at which the output resets.
9.2.2
Limit value monitoring OUT1 or OUT2 /
window function
▪
Select [SELx] and set FLOW
▪
Select [oux] and adjust the switching signal:
-
Fno = window function / normally open
-
Fnc = window function / normally closed
▪
Select [FHx] and set the value at which the output
switches.
▪
Select [FLx] and set the value at which the output resets.
48
Menu OUTx:
[SELx]
[oux]
[SPx]
[rPx]
Menu OUTx:
[SELx]
[oux]
[FHx]
[FLx]