Section 4-3
Functional Checks
4-3-1

Basic Controls

For a functional check of the system's features, including the Control Panel, Touch Panel, Monitor,
Keyboard and Trackball, refer to the LOGIQ™ S6 Quick Guide.
4-3-2

Performance Tests

4-3-2-1
Recommended Test Phantoms
GE Healthcare recommends the RMI 430GS phantom, but it is not required. It is the most current
phantom recommended to our field service personnel and provides the necessary targets and extended
life necessary for consistent system testing.
Resolution
Gray Scale
Photography
3
Lateral Distance
Measurement
Axial Distance
Contrast
Measurement
&
4
3
2
Resolution
RMI 403GS
Figure 4-28 Performance Test
Section 4-2 - General Procedure
Penetration
1
Lateral
Axial
Resolution
LOGIQ™ S6 S
ERVICE
2
4
6
8
Functional
Resolution
10
12
14
16
M
ANUAL
4-17