GE LOGIQ S6 Service Manual page 303

Hide thumbs Also See for LOGIQ S6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5162630, R
IRECTION
EVISION
10-7-6-3
Dale 600 Meter Procedure (cont'd)
5.) Set the meter's "FUNCTION" switch to LEAD TO GROUND position to measure the patient lead to
ground leakage current.
6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the
power condition combinations.
Table 10-19 Testing Power Conditions
3
Figure 10-1 ECG Leakage Current Test
ECG Power
Meter's Polarity Switch
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Section 10-7 - Electrical Safety Tests
ECG
Patient
Cable
Meter's Neutral Switch
NORM
NORM
REVERSE
REVERSE
NORM
NORM
REVERSE
REVERSE
LOGIQ™ S6 S
M
ERVICE
ANUAL
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
10-21