GE H
EALTHCARE
D
5162630, R
IRECTION
EVISION
8-3-8-6
TCP/IP Filter Setup (cont'd)
8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click OK buttons for all
opened window to complete the TCP/IP Filter Setup.
9.) The following dialog box appears. Click No to proceed the other settings. (After completion of all
system checks, the system is shut down.)
In case the last dialog box to restart the computer does NOT appear, check the TCP/IP Filter setup
NOTICE
again.
Connectivity check will be performed in
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
3
Figure 8-146 TCP/IP Filter Setup
Figure 8-147 Local Network
Section 1-13 Post Installation
LOGIQ™ S6 S
M
ERVICE
Setup.
ANUAL
8-31