EALTHCARE
D
5162630, R
IRECTION
EVISION
4-2-6

Archiving and Loading Presets for R6.x.x

NOTE:
Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an MO/
DVD-R disk for reloading on the system.
Presets should NOT be saved on the same MO/DVD-R disk as images. The Archive Menu lists the
NOTICE
images but does NOT list the presets stored on a /MO/DVD-R disk.
4-2-6-1
Formatting MO/DVD-R Disk
1.) Insert an empty (blank) MO/DVD-R disk into the MO/DVD-R device.
2.) Access to the Utility Menu on the Touch Panel, and select Connectivity>Tools. The Tools screen
will be shown on the monitor.
3.) Select the removable media from media list.
4.) Type a name for the removable media in Label field.
5.) Select Format button.