Required Features; Desirable Ultrasound Room Facilities - GE LOGIQ S6 Service Manual

Hide thumbs Also See for LOGIQ S6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5162630, R
IRECTION
EVISION
2-3-2

Required Features

Dedicated single branch power outlet of adequate amperage (see Table 2-10) meeting all local and
national codes which is located less than 2.5 m (8 ft.) from the unit's proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
NOTE:
The LOGIQ™ S6 has three outlets inside the unit for on board peripherals.
Power outlets for other medical equipment and gel warmer
Power outlets for test equipment and modem within 1 m (3.2 ft.) of unit
Clean and protected space to store transducers (in their cases or on a rack)
Material to safely clean probes (done with a plastic container, never metal)
2-3-3

Desirable Ultrasound Room Facilities

Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe sheaths
Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and service manual
2-8
3
Section 2-3 - Facility Needs
LOGIQ™ S6 S
M
ERVICE
ANUAL