IRECTION
EVISION
8-3-4-1
Processes prior to R7.7.x Installation
POWER
ON
*1: Loading R7.7.x application software procedures requires, at minimum, C: Drives to be ghosted (as
NOTICE
opposed to full ghost). In such case, patient image data, preset parameters, and customer data will be
maintained. Nonetheless, for safe-keeping customer data, it is recommended to back up all data prior
to R7.7.x installation.
In case of Full Ghost, customer data/setting must be saved for restoration after the process.
8-3-4-2
Processes to install R7.7.x
*2
*2: Do not power ON the peripherals when installing base system image. Disconnecting LAN is also
NOTICE
recommended procedure to ensure stable process.
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
3
Check SW
Power OFF
Level
USB Devices
Figure 8-113 Pre-installation Procedures
Figure 8-114 Software Loading Procedures
LOGIQ™ S6 S
Important Item
Caution Item
POWER
Save Customer
*1
Setting
M
ERVICE
ANUAL
OFF
8-7