After application is successfully installed, various LOGIQ™ S6 system parameters must be properly set.
Table 8-67 System Setting Summery
Item
Software Version
Computer Name
Patient
Database Check
LCD Monitor Setting
Software Option
VCR
Language pre-work (For
Japanese Only)
8-36
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
3
Method
Utility
\ System
\ About
Utility
\ Connectivity
\ TCP/IP
\ Computer Name
\ Enable DHCP
\ IP Address
\ Subnet Mask
\ Default Gateway
Utility
\ Connectivity
\ DataFlow
\ Default Data Flow
Press Patient (Keyboard or Touch Panel)
Press 'Toggle Button' for few seconds
Utility
\ Admin
\ System Admin
\ HW Number
Utility
\ Peripherals
\ Video Setting
\ VCR
Utility
\ Regional Option
\ Language (tab)
\ Install Files to East Asian Language
(Reboot)
LOGIQ™ S6 S
M
ERVICE
Proper Value
s/w version : R7.7.x
Image Part No. : 5305207
Same as Serial Number
OFF
restore customer setting
restore customer setting
restore customer setting
Local Archive – Int. HD
(n/a – Automatic)
(various)
Same as Option Dongle ID
PAL/NTSC-depends on region
(set accordingly)
ON / Apply