Modify System/Windows Settings
After Base System is successfully installed, various parameters must be properly set before proceeding to Application
load. The below outlines necessary system/windows setting.
Table 8-66 Summery of modification
Item
UPS Setting
(Restart System)
Avance AC97 Audio
(Restart System)
Drive Letter
Matrox for CRT
HW Device
LAN
TCP/IP Setting
Date/Time
8-16
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
3
Method
Automatic Setting
Load from c:\windows\system32\drivers
Automatic Setting but check desktop icon if not.
Check off automatic pop-up window
My Computer
\ Properties
Desktop
\ My Network Places
\ Properties
\ Local Area Connection
Desktop
\ My Network Places
\ Properties
\ Local Area Connection
\ Internet Protocol (TCP/IP)
\ Properties
\ Advanced
\ Options
\ TCP/IP Filter
Start
\ Settings
\ Control Panel
\ Date and Time
LOGIQ™ S6 S
M
ERVICE
Proper Value
(n/a – just load)
(n/a – just load)
DVD Drive = G
Set Taskbar into CRT Display
CRT Display mode: 800 x 600 75Hz
Touch Panel Display mode: 640 x
480 60Hz
Various
Double Click to activate
Values in 'Permit Only'
Proper Time Zone