Specifications; Package Contents; Installation; Unpacking The Centrifuge - Labnet Prism Instruction Manual

Microcentrifuge
Table of Contents

Advertisement

2
|
LABNET PRISM MICROCENTRIFUGE

3 .0 Specifications

Dimensions (W x D x H)
Maximum Speed
Maximum RCF
Maximum Volume
Timer
Admissible Sample Density
Electrical/Fuse Rating
Operating Conditions

4.0 Package Contents

Description
Prism™ Microcentrifuge
Standard angle rotor (Cat. No. 24D-RTR)
Rotor removal tool
Instruction manual
Warranty card
Power cord

5.0 Installation

The centrifuge should be installed on a rigid, even surface such as a stable laboratory bench,
countertop, etc. To guarantee sufficient ventilation, ensure that the centrifuge has at least 15 cm
(6 inches) of free space on all sides, including the rear. The centrifuge should not be located in areas
subject to excessive heat such as in direct sunlight or near radiators or the exhaust of a compressor, as
a buildup of heat may occur within the chamber.
Before operating the centrifuge, check that the power source (electrical outlet on the wall)
corresponds to that on the manufacturer's rating label, then connect the power cord to the centrifuge
and the power source.
10.25 x 13.8 x 7.48 in.
(26 x 35 x 19 cm)
15,000 rpm
21,200 x g
24 x 2.2 mL
0.5 to 99 minutes or continuous
1.2 kg/dm
3
120V, 50-60Hz, 1.9A, 5AT
230V, 50-60Hz, 1.1A, 2.5AT
0°C to 40°C / ≤80%RH
Quantity
1 each
1 each
1 each
1 each
1 each
1 each