5.3 Stop inhalation

1. Check the oxygen reserve at the contents
pressure gauge of the pressure reducer.
2. Close the oxygen supply valve and wait until the
system is depressurized.
3. Turn dial
to the left-hand stop (to "zero"). This
5
prevents the pressure gauge being damaged by
hammering when the device is switched back
on.
4. Switch the module off using toggle switch
5. Clean and disinfect/sterilize the device and the
device parts (see "11. Hygiene treatment" on
page 39).
6. Perform a function check (see "12. Function
check" on page 42).
Oxygen MODULE
.
4
23
EN