Weinmann Module System Instructions For Use Manual page 19

Table of Contents

Advertisement

2. Put both aspirate bottles in the holder.
3. Connect the bottles and the module with the
appropriate tubes. Ensure that the bottle with
the overflow cutout is connected to the module
first.
We recommend fitting a hydrophobic bacteria filter
WM 22291 for the exhaust air to prevent secretion
and bacteria penetrating the module. Filter
WM 22291 is obtainable as an accessory, individually
or in sets. It is a disposable item and must be replaced
after every aspiration operation.
4. Take the vacuum tube off the module.
5. Cut a piece about 6 cm long off the vacuum
tube.
6. Put this short piece of tube onto the side of the
filter marked with "TO PUMP".
7. Plug the other end of the short piece of tube
onto the vacuum connection of the module.
8. Put the free end of the vacuum tube on the filter.
Assembly
EN
19