Weinmann Module System Instructions For Use Manual page 33

Table of Contents

Advertisement

I
0
2,7 - 6 bar
O
2
hPa
Method
1. Switch on the oxygen supply to the CPAP valve
MODUL
by switching toggle switch
CPAP
2. If oxygen escapes from the O
off by turning dial
reach the stop.
3. Connect the CPAP valve to the CPAP MODULE
(see "4.7 Assembling the CPAP valve" on
page 20).
4. Connect the CPAP valve to the breathing mask.
5. Put the breathing mask on the patient.
6. If the patient tolerates the breathing mask,
attach it in position using the head band.
7. Increase CPAP pressure to a pressure of approx.
2 mbar by turning the regulator dial to the right.
8. Check that the breathing mask is properly
located so it does not leak. Correct the position
of the breathing mask if necessary.
9. Increase or reduce the flow of oxygen using the
regulator dial until the desired CPAP therapy
pressure is reached. To increase therapy
acceptance, increase pressure slowly (approx.
1 mbar/ 30 seconds).
10. Check the position of the breathing mask again
for leaks or for excessive pressure on the
patient's face. Correct the position of the
breathing mask if necessary.
11. Monitor the patient's therapy with the aid of the
displays (CPAP MODULE and gas cylinder
pressure gauges).
to
.
4
I
outlet, switch it
2
to the left until you feel it
5
CPAP MODULE
33
EN