AMI Codes-II CC
Program List and Explanations
3.3 Simulation
3.4 Set Time
3.5 Cleaning
3.5.1
3.5.1.1
3.5.1.20
A-96.250.581 / 181120
To simulate a value or a relay state, select the
alarm relay,
relay 1 or 2
signal output 1 or 2
valve 1 or 2
with the [
] or [
Press the [Enter] key.
Change the value or state of the selected item with the [
[
] key.
Press the [Enter] key.
The value is simulated by the relay/signal output.
Alarm Relay:
Relay 1 & 2:
Signal Output 1 & 2: Actual current in mA
Signal Output 3:
Valve 1 (photometer) Active or inactive
Valve 2 (KI supply)
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
Adjust date and time.
Automatic cleaning process using the optional Cleaning Module-II.
Cleaning is not possible if one of the following errors is active:
E009/E010 Sample flow high/low
E023 Cleaning solution
Parameters
Mode: The following modes can be chosen: interval, daily, weekly or
off.
If Mode = Interval
Interval: Select one of the following cleaning intervals:
1 h, 2 h, 3 h, 4 h, 6 h, 8 h, 12 h.
] key.
Active or inactive
Active or inactive
Actual current in mA (if option is installed)
Active or inactive
] or
87