AMI Codes-II CC
Maintenance
Prepare
Oxycon
On-line Buffer
Prepare
Oxycon
On-line KI
Start-up
52
7 Remove the screw cover and insert the suction lance [A] and
tighten the screw cover.
1 Rinse the canister [F] labelled "OXYCON ON LINE Buffer" solu-
tion with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of buffer Oxycon On-line Buf-
fer into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [F] into holder [H].
7 Remove the screw cover, insert the suction lance [B] and tighten
the screw cover.
1 Rinse the canister [G] labelled "OXYCON ON LINE KI" solution
with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of Oxycon On-line KI into the
canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [G] into holder [H].
7 Remove the screw cover, insert suction lance [B] and tighten the
screw cover.
Fill reagent system. See
Fill or Flush Reagent System, p. 41
A-96.250.581 / 181120