AMI Codes-II CC
Maintenance
6.3.
Reagent
consumption
Measuring interval
fc
2 minutes
10 minutes
4 minutes
20 minutes
A-96.250.581 / 181120
Refill or replace Reagents
The liquid level in the containers 2 and 3 is monitored.
The following messages are displayed:
Container almost
empty
Container empty
Note: Before refilling the reagents, rinse the containers with
demineralized water.
The 2 liter reagent container for DPD & Buffer will last for 19 days of
operation (with measurement interval of 4 minutes). The 2 liter re-
agent container for KI will last for 66 days of operation (with mea-
surement interval of 20 minutes). The provided reagent set (for 8
containers) therefore lasts for 5 respectively 17 months of operation.
As the reagent consumption is not linear find further examples be-
low:
Duration of DPD/Buffer
one canister
tc
filling
~ 14 days
~ 19 days
WARNING
Health hazard
Some reagents can cause severe burns or eye damage.
For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS), see
Safety Data sheets, p. 106
Maintenance E065 - Reagents low and the
remaining reagent volume in %
(starting at 17 % = 340 ml).
Error E022 - Reagent empty
reagent set one canister
3 months
5 months
Duration of KI
reagent set
filling
~ 33 days
8 months
~ 66 days
17 months
Material
49