AMI Codes-II CC
Default Values
A-96.250.581 / 181120
Parameter: .................................................................Calc. monochl.
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ........................... 5.00 ppm
Settings: Control Parameters: P-band:............................. 0.10 ppm
Parameter: ................................................................. Calc. comb. cl.
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ........................... 5.00 ppm
Settings: Control Parameters: P-band:............................. 0.10 ppm
Parameter: .......................................................... Calc. dichloramine
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ........................... 5.00 ppm
Settings: Control Parameters: P-band:............................. 0.10 ppm
Parameter: .....................................................................Temperature
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ................................. 30 °C
Settings: Control Parameters: P-band:..................................... 1 °C
Parameter: .................................................................... Sample Flow
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ............................. 200 B/s
Settings: Control Parameters: P-band:................................. 20 B/s
Parameter: ..................................................................................... pH
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency: ................................................120/min
Settings: Control Parameters: Setpoint: ...................................7 pH
Settings: Control Parameters: P-band:..................................0.1 pH
Common settings:
Settings: Control Parameters: Reset time: ................................. 0 s
Settings: Control Parameters: Derivative Time: ......................... 0 s
Settings: Control Parameters: Control Timeout:..................... 0 min
Settings: Actuator:................................................... Time proportional
Cycle time: ............................................................................... 60 s
Response time: ....................................................................... 10 s
Settings: Actuator............................................................. Motor valve
Run time: ................................................................................. 60 s
Neutral zone: ............................................................................. 5%
109