D.3

Internal Shutter Installation Procedure

Perform the following procedure to install an internal shutter into the IsoPlane 81:
1.
Carefully slide the shutter into the bottom of the IsoPlane 81 chassis.
2. Carefully reconnect the two-conductor electrical shutter cable.
3. Use the 3/8" Allen wrench to secure/tighten the two Allen screws.
4. Replace the shutter access panel.
5. Replace the four (4) Phillips head screws to secure the shutter access panel to the
underside of the IsoPlane 81 chassis and use the #1 Phillips head screwdriver to
tighten each screw. See Figure D-2.
Figure D-2: Install Shutter Access Panel: Underside of IsoPlane 81
®