Install Additional Cubes - Teledyne Princeton Instruments IsoPlane 81 Manual

Imaging spectrometer system
Table of Contents

Advertisement

Chapter 4
4.4.1

Install Additional CUBES

If desired, additional CUBES may be stacked onto the Focusing CUBE. Figure 4-9
illustrates a typical configuration with three CUBES installed. Refer to Chapter 2 for a list
of compatible CUBES.
Figure 4-9: Typical IsoPlane 81 System with Three (3) CUBES Installed
Perform the following procedure to install additional CUBES onto the input to the
IsoPlane 81:
1.
Use the #1 Phillips head screwdriver to tighten the bottom screw on the previously
installed CUBE.
2. Install the second CUBE by aligning the stainless steel pins and carefully pressing
the CUBE onto the existing CUBE.
3. Using the #1 Phillips head screwdriver, turn the coupling screw clockwise until
finger tight.
4. Repeat step 1 through step 3 for each additional CUBE that is to be installed on the
IsoPlane 81.
Getting Started
NOTE:
The installed lens is not intended for wide-field imaging. A
diffraction-limited image will not be acquired. A camera
lens is necessary when performing wide-field imaging.
27

Advertisement

Table of Contents
loading

Table of Contents