22
4.3.1

Required Tools

The following tools are required to focus a slit:
4.3.2

Focus Procedure

Perform the following procedure to focus the slit:
1.
Illuminate the slit with light from the IsoPlane 81 atomic emission wavelength
calibration lamp.
2. If necessary, within LightField, on the Experiment Setting tab, open the Regions of
Interest expander and select Bin Center 10 rows. See Figure 4-1.
Figure 4-1: Configure Regions of Interest
3. On the Common Acquisition Settings expander, configure an Exposure Time of
50 ms. See Figure 4-2.
®
IsoPlane
81 Imaging Spectrometer System Manual
5/64 Allen wrench
#1 Phillips head screwdriver
NOTE:
If you do not have an IsoPlane 81 atomic emission
wavelength calibration lamp, use a suitable atomic emission
lamp.
Issue 1