Appendix A
A.3.1

Ventilation

A minimum of 1 inch (2.54 cm) clearance is required around all vents on the IsoPlane 81
system.
When the IsoPlane 81 is operated within an enclosure, >30 cfm air circulation and heat
dissipation of 200 W is required.
A.4

Cooling Specifications

Refer to Table A-4 for cooling specifications for the IsoPlane 81 system.
Table A-4: Temperature Specifications
Cooling Temperature
Temperature Stability
a. Guaranteed at ambient temperature of +20°C
A.5

CCD Specifications

Refer to Table A-5 for CCD specifications for the IsoPlane 81 system.
Table A-5:
CCD
Image Type
Resolution
Pixel Size
Active Area
Frame Rate
a. Specifications are valid as of the publication date
Technical Specifications
Specification
CCD Specifications
Parameter
of this manual. For up-to-date specifications,
refer to the IsoPlane 81 data sheets available for
download from www.princetoninstruments.com.
Cooling Temperature
Minimum
Default
a
-55
±1
a
Specification
Proprietary
Monochrome
1024 x 256
μ
μ
13
m x 13
m
13.3 mm x 3.3 mm
34 fps (full frame)
35
Units
Maximum
-60
°C
°C