6.4

Resource block

Error conditions found in the Resource block.
Table 6-6: Resource Block BLOCK_ERR Messages
Condition Name
Block configuration error
Simulate active
Power up
Out of Service
Table 6-7: Resource Block DETAILED_STATUS Messages
Condition Name
Sensor Transducer block error
Manufacturing block error
Non-volatile memory error
ROM integrity error
6.5

Transducer block

Error conditions found in the transducer block.
Table 6-8: Transducer Block BLOCK_ERR Messages
Condition Name
Other error
Out of Service
236
Description
Configuration Error is used to indicate that you have selected an
item in FEATURES_SEL or CYCLE_SEL that was not set in
FEATURES or CYCLE_TYPE, respectively
This indicates that the simulation switch is in place. This is not an
indication that the I/O blocks are using simulated data
The actual mode is out of service
Recommended action
1. Restart processor
2. Call service center
1. Restart processor
2. Call service center
1. Restart processor
2. Call service center
1. Restart processor
2. Call service center
Description
Set whenever XD_ERROR is non-zero. See also
status in AMS Device
Manager.
The actual mode is out of service.