Emerson Rosemount 5900C Reference Manual page 214

Radar level gauge
Table of Contents

Advertisement

Service and troubleshooting
September 2019
6.2.3
Recover a backup configuration database using
TankMaster
Rosemount TankMaster WinSetup lets you replace the current Holding Register database
with a backup database stored on disk. This can be useful, for example, if you want to
recover lost configuration data.
To load a Holding Register database do the following:
Procedure
1. In the TankMaster WinSetup workspace window, select the device icon.
2. Right-click and select Upload Database, or from the Service menu choose Devices/
3. Click the Browse button and choose a database file to be uploaded, or type a path
4. Click the Upload button.
6.2.4
View and configure diagnostic registers using
TankMaster
The Rosemount TankMaster WinSetup program lets you view the current device status.
The View Diagnostic Register window shows a selection of database registers that gives
214
Upload Database.
and file name.