Overview
September 2019
There are two versions of the Rosemount 2410 Tank Hub; one for single tank operation
and one for multiple tanks operation. The multiple tanks version of the Rosemount 2410
supports up to 10 tanks and 16 devices. With the Rosemount 5300 and Rosemount 5400
level transmitters the Rosemount 2410 supports up to 5 tanks.
The Rosemount 2410 is equipped with two relays which support configuration of up to 10
"virtual" relay functions allowing you to specify several source signals for each relay.
The Rosemount 2410 supports Intrinsically Safe (IS) and Non-Intrinsically Safe (Non-IS)
analog 4-20 mA inputs/outputs. By connecting an Emerson Wireless 775 THUM Adapter
to the IS HART 4-20 mA output, the tank hub is capable of wireless communication with an
Emerson Wireless Gateway in a WirelessHART
2.4.4
Rosemount 5900C Radar Level Gauge
The Rosemount 5900C Radar Level Gauge is an intelligent instrument for measuring the
product level inside a tank. Different antennas can be used to meet the requirements of
different applications. The Rosemount 5900C can measure the level of almost any
product, including bitumen, crude oil, refined products, aggressive chemicals, LPG, and
LNG.
The Rosemount 5900C sends microwaves towards the surface of the product in the tank.
The level is calculated based on the echo from the surface. No part of the 5900C is in
actual contact with the product in the tank, and the antenna is the only part of the gauge
that is exposed to the tank atmosphere.
2.4.5
Rosemount 5300 Guided Wave Radar
The Rosemount 5300 is a premium 2-wire guided wave radar for level measurements on
liquids, to be used in a wide range of medium accuracy applications under various tank
conditions. Rosemount 5300 includes the Rosemount 5301 for liquid level measurements
and the Rosemount 5302 for liquid level and interface measurements.
2.4.6
Rosemount 5400 Radar Level Transmitter
The Rosemount 5400 is a reliable 2-wire non-contact radar level transmitter for liquids, to
be used in a wide range of medium accuracy applications under various tank conditions.
2.4.7
Rosemount 2240S Multi-Input Temperature Transmitter
The Rosemount 2240S Multi-input Temperature Transmitter can connect up to 16
temperature spot sensors and an integrated water level sensor.
2.4.8
Rosemount 2230 Graphical Field Display
The Rosemount 2230 Graphical Field Display presents inventory tank gauging data such as
level, temperature, and pressure. The four softkeys allow you to navigate through the
different menus to provide all tank data, directly in the field. The Rosemount 2230
supports up to 10 tanks. Up to three Rosemount 2230 displays can be used on a single
tank.