Service
Conditions:
1 Skin temperature display range is called up.
2 Simulation resistances are used instead of the
temperature sensors.
The following basic values may be used:
33.2 °C (R= 1586 Ω)
36.0 °C (R= 1412 Ω)

Maintenance

WARNING
Risk due to faulty components
Device malfunction can occur due to wear and
material fatigue of components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
WARNING
Risk of electric shock
Before performing any service work,
disconnect all electrical connections and gas
connections from power and gas supplies.
Only perform service activities when there is
no patient in the device.
CAUTION
Risk of electric shock
Do not set up or store the device in a dirty
environment.
Check the device at regular intervals for dust
residues.
98
37.8 °C (R= 1312 Ω)
Procedure:
 Check the display for 3 different temperature
values.
Result:
 The displayed measured value may differ by no
more than 0.1 °C from the value simulated by
the respective resistance.
WARNING
Risk due to faulty heating elements
The heating elements are designed for safe
operation within the maintenance intervals. If
the maintenance intervals are not adhered to
or if the device is used not in accordance with
the specified intended use, the patient may be
at risk.
Comply with the specified maintenance
interval.
Instructions for use Babytherm 8004/8010