Operation

Using the bed canopy

The infant's core temperature should be
monitored when the bed canopy is in use.
CAUTION
The radiant warmer is less effective when the bed
canopy is used. The patient's core temperature
should be monitored in order to prevent adverse
effects on the temperature balance.
NOTE
Use of the bed canopy in conjunction with
operation of the integrated phototherapy system
leads to a significant reduction in the bilirubin-
effective radiant intensity and a very uneven
distribution of the radiation on the bed.
WARNING
When the bed canopy is in place, the side
panels can be folded down. Do not leave the
Babytherm unattended with the side panels
open, as the infant may fall out.

Installing the bed canopy

Check whether the bed canopy is in the closed
position; otherwise:
58
1 Close bed canopy = turn lock until it engages.
2
2 Hold bed canopy handles with both hands and
place over the side panels.
It is advisable to use the higher side panels
(230 mm) when using the bed canopy.

Opening/closing the bed canopy

1
1 Unlock bed canopy = turn lock
2 Open bed canopy = swivel handles upwards
until they engage. The infant is now accessible
for care/medical treatment.
2 Close bed canopy = swivel handles forwards.
Instructions for use Babytherm 8004/8010