Show Operating Hours For Phototherapy; Test Buttons On Control Panel - Dräger Babytherm 8004 Instructions For Use Manual

Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Configuration mode

Show operating hours for phototherapy

2
3
3
1
1 Briefly press the
2 The following message appears on the display:
Configuration Mode
Mode c04 : Time counter photo therapy.
3 Indication of the operating hours up to
1000 hours.
4 Press
button for 3 seconds. The counter is
reset to 0. This is necessary when changing the
phototherapy lights.

Test buttons on control panel

2
3
4
skin C
manual
skin
man .
1
1 Briefly press the
2 The following message appears on the display:
Configuration Mode
Mode c05 : Keyboard test (inactive).
74
Babytherm 8010
button.
Babytherm 8010
10
15 min
C
5
1
< 36 C
mattress . C
5
button.
3 The abbreviation for this mode appears: but
4 The code for the current mode appears: c05
5 Press the
2 The following message appears on the display:
4
Configuration Mode
Mode c05 : Keyboard test (active).
Button numbers
1
1
3
2
8
9 10
10
 Press the buttons to be tested,
1 A string of digits appears on the display:
1st digit = number of buttons currently pressed,
2nd and 3rd digit = number of last button
pressed.
 Press the
ended.
Instructions for use Babytherm 8004/8010
or
button to activate the test.
Babytherm 8010
4
5
6 7
13
13
11 11
12
12
button for 2 seconds. The test is
15
15
14
14

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents