Dräger Babytherm 8004 Instructions For Use Manual page 46

Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Operation
Physiological dangers of oxygen
CAUTION
Only enrich the O
concentration under the
2
instructions of a doctor.
Only do this in accordance with the arterially
measured O
partial pressure in the patient's
2
blood. Otherwise there is a danger of
hyperoxemia (which may cause damage to the
eyes) or hypoxemia (which may cause damage to
the brain). During oxygen therapy, the oxygen
concentration must be constantly monitored, e.g.
with MiniOx 3000.
X-ray tray
CAUTION
Do not use the X-ray tray in the extended position
to rest on while writing!
Do not place any objects on the tray!
Do not lean on the tray!
Risk of damage!
NOTE
Remove X-ray cartridge from the X-ray tray during
transport.
Bed canopy
The infant's core temperature should be
monitored when the bed canopy is in use.
CAUTION
The radiant warmer is less effective when the bed
canopy is used. The patient's core temperature
should be monitored in order to prevent adverse
effects on the temperature balance.
46
WARNING
Risk of excessive exposure to radiation
Do not place a phototherapy light on the bed
canopy. The distance to the patient is too
small.
Observe the minimum distance of the
phototherapy light (see the corresponding
instructions for use).
NOTE
Use of the bed canopy in conjunction with
operation of the integrated phototherapy system
leads to a significant reduction in the bilirubin-
effective radiant intensity and a very uneven
distribution of the radiation on the bed.
NOTE
Higher side panels (230 mm) are recommended
when using the canopy.
WARNING
When the bed canopy is in place, the side
panels can be folded down. Do not leave the
Babytherm unattended with the side panels
open, as the infant may fall out.
NOTE
The bed canopy should not be used as a
convenient place to lay objects, clothing, etc.
Configuration mode
Do not perform care/therapy during configuration
of the device, because the normal equipment
functions (e.g., temperature measurement, alarms)
are switched off.