O
therapy
2
CAUTION
Only enrich the O
concentration under the
2
instructions of a doctor. Only do this in
accordance with the arterially measured O
partial pressure in the patient's blood. Otherwise
there is a danger of hyperoxemia (which may
cause damage to the eyes) or hypoxemia (which
may cause damage to the brain).
Monitor the O
concentration during the O
2
therapy.
Injector and closed bed canopy
6
7
Place canopy in position (see page 58).
9 Set O
concentration on the injector: 30, 40, 50
2
Vol.% O
.
2
Set O
supply on flowmeter.
2
Injector setting
Vol.% O
O
supply
L/min
2
10 Monitor the O
concentration.
2
After use
Open the bed canopy.
Close the valve on the O
clockwise.
Instructions for use Babytherm 8004/8010
2
2
30
40
50
2
5
10
15
flowmeter = turn
2
Dismantle and reprocess the therapy
equipment, see page 82.
Bronchial aspiration
Use bronchial aspirator in accordance with its
specific Instructions for Use.
Configuration mode
77