Possible Error Messages - Hart® Devices From Sw-Rev. 3.00 - ABB TTH300 Operating Instruction

Head-mount temperature transmitter
Hide thumbs Also See for TTH300:
Table of Contents

Advertisement

70
TTH300 HEAD-MOUNT TEMPERATURE TRANSMITTER | OI/TTH300-EN REV. G
... 11 Diagnosis / error messages
Possible error messages – HART® devices from SW-Rev. 3.00
Note
The listed causes for a device status message correspond to the delivery status. They can be configured freely in the tools in the
menu "Diagnosis/NAMUR configuration", see interface description COM/TTX300/HART.
Range
Device status message
(on the display)
Sensor
Line resistance S1 too high
Sensor
S1 short-circuit
Sensor
S1 Wire break / sensor break
Sensor
S1 Single-point trim is active
Sensor
S1 Two-point trim is active
Sensor
Line resistance S2 too high
Sensor
S2 short-circuit
Sensor
S2 Wire break / sensor break
Sensor
S2 Single-point trim active
Sensor
S2 Two-point trim active
Sensor
Redundancy: S1 not available
Sensor
Redundancy: S2 not available
Sensor
Sensor drift detected
Sensor
Application failure
Operation
Diagnostics is simulated
Operation
Analog output fixed / simulated
Operation
Application warning
Electronics
Electronics temp. measurement failure
Electronics
Electronics temp. out of spec.
Electronics
Non-volatile data defect
Electronics
Non-vol. Memory Write cycles exceeded
Electronics
Device not calibrated
Electronics
Electronics failure
Process
S1 over sensor range
Process
S1 under sensor range
Process
S2 over sensor range
Process
S2 under sensor range
Process
High limit: Alarm
Process
Low limit: Alarm
Process
High limit: Warning
Process
Low limit: Warning
Configuration
Parameterization / config. failure
Note
If the remedial measures listed for the error message do not improve the status of the device, please consult ABB Service.
Cause
Remedy
Maintenance required
Check sensor or replace / repair sensor
Maintenance required
Check sensor or replace / repair sensor
Maintenance required
Check sensor or replace / repair sensor
Check function
Check function
Maintenance required
Check sensor or replace / repair sensor
Maintenance required
Check sensor or replace / repair sensor
Maintenance required
Check sensor or replace / repair sensor
Check function
Check function
Maintenance required
Check sensor or replace / repair sensor S1
Maintenance required
Check sensor or replace / repair sensor S2
Maintenance required
Check Sensor/ connection / trim / drift parameter
Error
Check sensor connection / check HART variable
mapping
Check function
Terminate / come out of simulation
Check function
Terminate / come out of simulation
Check function
Load valid parameter restart (reset) check S1 / S2
Error
Restart (RESET) or replace transmitter
Out of specification
Observe spec. ambient temp. range
Error
Restart (RESET) or replace transmitter
Maintenance required
Restart (RESET) or replace transmitter
Error
Restart (RESET) or replace transmitter
Error
Restart (RESET) or replace transmitter
Maintenance required
Check sensortype use diff. sensor if required
Maintenance required
Check sensortype use diff. sensor if required
Maintenance required
Check sensortype use diff. sensor if required
Maintenance required
Check sensortype use diff. sensor if required
Out of specification
Verify process or change limit value
Out of specification
Verify process or change limit value
Out of specification
Verify process or change limit value
Out of specification
Verify process or change limit value
Error
Load valid param restart (RESET), reset to factory