TTH300 HEAD-MOUNT TEMPERATURE TRANSMITTER | OI/TTH300-EN REV. G
Range
Displays the
device
status
Sensor
M
Sensor
M
Sensor
M
Installation / Configuration
C
Operating conditions
F
Installation / Configuration
I
Installation / Configuration
C
Operating conditions
S
Operating conditions
S
Operating conditions
S
Operating conditions
S
Operating conditions
S
Note
If the remedial measures listed for the error message do not improve the status of the device, please consult ABB Service.
Displays the
Cause
DIAG. NO.
66
No sensor 1 detected during redundancy
configuration
67
No sensor 2 detected during redundancy
configuration
68
Sensors exceeded specified drift window
71
Reconfiguration is running
72
Error in the application
74
Analog output adjustment active
75
Analog output in simulation
76
Values overshot
77
Limit HIGH HIGH
78
Limit LOW LOW
79
Limit HIGH
80
Limit LOW
Remedy
Check connection
Check connection
Calibrate sensors
Terminate reconfiguration
Check configuration, connections; reset to factory
settings
Terminate compensation
Terminate simulation
Check parameters:
A) Sensor limits overshot
B) Span is too small
Upper limit value: alarm
Lower limit value: alarm
Upper limit value: warning
Lower limit value: warning