Appendix J: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Nanoduct system is intended for use in the electromagnetic environment specified below. The customer or the
operator of the Nanoduct system should assure that it is used in such an environment.
IEC 60601-1-2 Test Level
Immunity
Frequency
Test
(MHz)
800 – 2700 10
385
450
710
745
Radiated
780
RF
Immunity
810
IEC
870
61000-4-3
930
1720
1845
1970
2450
5240
5500
5785
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
a
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Nanoduct system is used exceeds
the applicable RF compliance level above, the Nanoduct system should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
Nanoduct system.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
Level 
Frequency
Modulation
(V/m)
(MHz)
1 KHz 80%
Amplitude
800 – 2700
Modulation
Pulse
27
Modulation
385
18 Hz
FM ± 5 Hz
deviation
28
450
1 KHz sine
710
Pulse
9
Modulation
745
217 Hz
780
810
Pulse
28
Modulation
870
18 Hz
930
1720
Pulse
28
Modulation
1845
217 Hz
1970
Pulse
28
Modulation
2450
217 Hz
5240
Pulse
9
Modulation
5500
217 Hz
5785
Compliance Level
Level 
Modulation
(V/m)
1 KHz 80%
3
Amplitude
Modulation
Pulse
27
Modulation
18 Hz
FM ± 5 Hz
deviation
28
1 KHz sine
Pulse
9
Modulation
217 Hz
Pulse
28
Modulation
18 Hz
Pulse
28
Modulation
217 Hz
Pulse
28
Modulation
217 Hz
Pulse
9
Modulation
217 Hz
Electromagnetic
Environment - Guidance
Recommended minimum
separation distance (m)
1.2√P (80 – 800 MHz)
2.3√P (800 MHz – 2.7 GHz).
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey a, should be less than
the compliance level in each
frequency range. b
Interference may occur in the
vicinity of equipment marked
with the following symbol: