Elitech NANODUCT 1030 User Manual page 44

Neonatal sweat analysis system
Table of Contents

Advertisement

Section 4: Troubleshooting and Preventive Maintenance
Error Message
Displayed
Open Loop
Fault
Check Electrodes
Screen
Iontophoresis
Probable Causes:
Possible Solutions:
If fault occurs on a regular basis, contact ELITechGroup for further
instructions.
Probable Causes / Possible Solutions
No Pilogel discs under electrodes.
Dirty electrodes.
Electrodes not properly latched with locking ring or cable
not connected to Nanoduct.
Holders not adequately secured to patient.
Damaged electrode cable assembly.
Ensure that a Pilogel disc is under each electrode.
Clean electrodes if necessary. Place a drop of deionized
water directly on the clean skin beneath each Pilogel disc.
Ensure both electrodes are latched with locking ring and
that the cable is properly connected to the Nanoduct.
Ensure holders are secured with adequate tightness to
the patient's limb. Do not overtighten and impede blood
circulation.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires.
Repeat the sweat test one time.†
44

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents